-40%

Contec MS400 Multi-parameter Patient Simulator,ECG Simulator USA Warehouse,Fedex

$ 210.67

Availability: 82 in stock
  • Refund will be given as: Money Back
  • Country/Region of Manufacture: China
  • Model: MS400
  • All returns accepted: Returns Accepted
  • MPN: 69450401
  • Condition: New
  • Return shipping will be paid by: Buyer
  • Brand: CON-TEC
  • Item must be returned within: 60 Days

    Description

    USA Stock Contec New Brand MS400 Multi-parameter Patient Simulator,ECG Simulator
    Introduction
    MS400 Multiparameter Simulator is a portable device, features in  small volume and light weight. It can simulate accurately 12-lead ECG,  Respiration, Temperature, 2-channel IBP or 4-channel IBP. And it can be tested  in where the monitor is used.
    Features
    Simulate 35 kinds of arrhythmia (including Atrial, Ventricular,  Premature Beat and Conduction Defects) and pacemaker waveforms (including  Asynchronous, Demand Occ Sinus, Demand Freq Sinus, A/V Sequential, Non Capture,  Non Function).
    Simulate four-channel IBP data.
    Simulate respiration signal with adjustable respiration rate, impedance and  baseline impedance, and three kinds of apnea data.
    Simulate four kinds of temperature data.
    Simulate square wave, triangle wave, pulse wave and sine wave with  adjustable frequency.
    Simulate standard ECG calibration, adult and pediatric ECG signal.
    ECG waveform complied with the standard of ANSI/AAMI EC13-2002, AHA, MIT ECG  database, other ECG database and user-defined waveform data stored in TF card  can be displayed by the simulator and exported to the equipments to be  tested.
    Each waveform can be displayed by sliding the 3.5'' TFT touch screen, easy  to operate.
    Power supply: rechargeable lithium battery or linear adapter, which reduces  interference.
    Optional operation modes: keys and touch screen.
    Parameters:
    ECG
    Conventional: All leads have the same voltage with  the II-lead as the reference
    Amplitude: 0.5,1.0,1.5 and 2.0mV
    ECG Rate:  15,20,25,30,40,60,80,100,120,140,160,180
    200,220,240,260,280,300,320,340 and 350  BPM
    ECG artifacts: 50 and 60 Hz, muscle, baseline and breathing
    The elevation of the ST segment: +0.8 to -0.8 mV, the step is 0.1 mV
    Arrhythmia
    Atrial: Atrial Fib(coarse), Atrial fib(fine),  Atrial Flutter, Sinus Arrhythmia, Missed Beat, Atrial Tachycardia, Nodal Rhythm,  Supraventricular Tachycardia
    Ventricular: Vent Fib (coarse), Vent Fib (fine),  PVCs 6/min, PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy,  Paired PVCs, Run 5 PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
    Premature  Beat: Atrial PAC, Nodal PNC, PVC1 LV Focus, PVC1 E LV Focus, R on T LV Focus,  PVC2 RV Focus, PVC2 E RV Focus, R on T RV Focus, Multifocal PVC
    Conduction  Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart  Block, Right Bundle Branch Block, Left Bundle Branch Block
    Four IBP Channels
    The default: 00/00mmHg
    Dynamic  IBP(unit: mmHg)
    Channel 1:
    Artery: 120/80
    Radial: 120/80
    Left Ventricle: 120/00
    Right  Ventricle: 25/00
    Channel 2:
    Artery: 120/80
    Radial: 120/80
    Left Ventricle: 120/00
    Right  Atrium(CVP): 15/10
    Right Ventricle: 25/00
    Pulmonary Artery:  25/10
    Pulmonary Wedge: 10/2
    Left Atrium: 14/4
    Channel 3:
    Artery: 120/80
    Radial: 120/80
    Left Ventricle: 120/00
    Right  Atrium(CVP): 15/10
    Right Ventricle: 25/00
    Pulmonary Artery:  25/10
    Pulmonary Wedge: 10/2
    Left Atrium: 14/4
    Channel 4:
    Auto  Swan-Ganz (every 15 seconds): RA(CVP), RV, PA, PAW (Pulmonary Wedge)
    Man  Swan-Ganz: press ENTER key to change every time
    Static-Pressure  Levels
    Channel 1: 0, 80, 160, 240, 320 and 400mmHg
    Channel 2: 0, 50, 100,  150, 200 and 250mmHg
    Channel 3: 0, 20, 40, 60, 80 and 100mmHg
    Channel 4:  P4 tracks the static pressure levels of P3
    Respiration
    Respiration Rate: 0 (Apnea), 15, 20, 30, 40,  60, 80, 100, 120 and 150 R/MIN
    Impedance: 0.2, 0.5, 1.0 and 3.0 Ω
    Baseline  impedance: 500, 1000, 1500, 2000Ω
    Apnea: 12 s, 22 s, 32 s
    Temperature
    Four temperature: 0℃(32°F), 24℃( 75°F),  37℃(98.6°F), 40℃(104°F)
    Accessories
    Standard accessories 9V/1A(CE) power adapter, an User manual, a  packaging box, a plastic packaging.
    Optional accessories a power  adapter(15V/1A)(UL), a power adapter(15V/1A), a temperature cable, an IBP cable,  a CD disk(including a case conversion software)
    Physical  characteristic
    Dimension 18.86 cm(L)×12.50 cm(W)×4.32 cm(H)
    Net  weight about 1.4kg
    Buy Safe Product!
    The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE:
    The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA regulation, We will verify your status as an authorized purchaser of this item before shipping of the item.
    Shipment
    1. We will ship the item within 48 hours after receiving your payment, please kindly leave your detail shipping address and telephone number, both are very important for the shipment,thank you.
    2.We will upload the tracking number after shipment,you can search from both ebay and paypal.You can track online.
    3. If you haven't received the item after 2 weeks of your payment, please contact me immediately!
    Terms of Sale
    1.If you have a defective item, you want to return or discount. Please contact us within 60 days from you receive the shipment.All return items must be returned with it's original packaging and accessories. Customer is responsible for shipping charges on returned items.
    2.we will refund the money to you when we get the return items. Or replace item for you. We will deal with the refund in 2 business days when we get the info from you.Please read our trade Policy before buy , if you don't agree our trade policy please don't buy it.
    Only English user manual,if you need any other language,please contact us!