-40%
Bionet Cardiocare 10 Lead 12 Ch. Interpretive ECG-EKG Cardiocare 2000
$ 662.64
- Description
- Size Guide
Description
Bionet Cardiocare 10 Lead 12 Ch. Interpretive ECG-EKG Cardiocare 2000The CardioCare 2000 is an economically priced 12 channel resting ECG that does not compromise on performance. Suitable for use in private practice, the ER, or hospitals use, the CardioCare 2000 is designed for maximum ease of use and convenience. Software upgradeable and networkable, the CardioCare 2000 is an excellent choice for your diagnostic needs.
Simultaneous 12 Channel Resting ECG
The CardioCare 2000 acquires 10 full seconds of ECG data simultaneously from 10 leads to produce an accurate 12 channel printout with analysis!
ECG Interpretation Program
Over 80 kinds of interpretation results based on the advanced Minnesota Code are provided to aid your diagnosis. The interpretation algorithms are expanded with regular software updates.
A4 Full Size Paper
The CardioCare 2000 will produce full size A4 printouts in 1,3,6 and 12 channel formats. Full size print-outs improve readability and make it easier for filing in the patient chart. A convenient Copy button lets you generate ECG copies in any of the available formats.
Low Maintenance Cost
The CardioCare 2000 uses normal fax paper which costs about ¼ the cost of ECG paper. It will even generate its own grid!
Specifications:
ECG Leads
10 lead 12 channel simulataneous 10-second resting ECG
Dimension
296 (w) x 305.5 (h) x 92.5 (d) mm Approximately 2.98 kg
Recording Channel
3,6,12 channel and 1 channel (60 sec)
Sensitivity
5, 10, 20 auto (I~aVF: 10, V1~V6:5) mm/mV
Printing Speed
12.5, 25, 50 mm/s
Filters
AC (50/60 Hz, -20dB or better)
Muscle (25~35Hz, -3dB or better)
Base Line Drift (0.1Hz, -3dB or better)
Low pass filter (off, 40Hz, 100Hz, 150Hz)
Display
2 x 16 character LCD display
LCD display
Sensitivity, speed, filter status, HR, printing form, rhythm lead
LED indicator
Signal quality, power source, battery
Keyboard
Membrane keyboard Alphanumeric and symbol available
Patient Data
ID, name, age, sex, height, weight
Basic Measurement
Heart rate, PR, QRS, QT/QTc, P-R-T axis
Recorder
Thermal print head, thermal roll paper
Report papers:
-Width: A4: 210mm or 8.5"
-Length: A4: 300mm or 11"
Effective writing width: 204mm
Resolution:
- Vertical: 8 dots/mm
- Horizontal: 16 dots/mm
Electrical
Internal noise: 20uV(p-p) max
Input circuit: Floating input
Isolated and defibrillation protected
Input impedance: >10MQ
Input range: > +/- 5mV
Common Mode Rejection: > +100dB
DC offset voltage: > +/- 300mV
Time constant: 3.2 sec
Patient leakage current: <10uA
Frequency response: 0.05~150Hz
Signal qualilty control
Disconnected lead detection
Power
Power Supply: AC or built-in Battery (option) Power requirement: 100 - 240 Vac, 50/60Hz, 1.0 - 0.5A Power consumption: 60w max
Battery capacity
1 hour of normal use (approximately 100 automatic ECG printouts)
Communication
PC connection with RS-232 interface and LAN
Safety
Class 1, Type BF
Conformity
CE, CSA
Environment
Operating humidity: 30~85% Operating temperature: 10° ~40° C Atmospheric pressure: 70 ~ 106 Ka
What does it come with?
CardioCare 2000 Main Unit
10 Lead ECG Patient Cable
Alligator Clips
Electrodes
Medi-Graph Paper
Hospital Grade Power Cord
Operation Manual
Diagnostics Guide Book
Quick Reference Guide
4-Year Warranty Card.
"The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item."